Summary
Type: SPE
Application #: 300930
Catalog#: 730001 - SPE Cartridges, Chromabond, C18, 1ml, PP, 100mg, 100/pk
Detection: HPLC
Conditions: Column conditioning: with 2 x 1 ml of MeOH followed by 2 x 1 ml of 0,5 % AcOH
Sample application: A 160-μl volume of the resulting solution was loaded on the cartridge in duplicate and drawn trough completely under light vacuum.
Column washing: The cartridges was allowed to stand for 1 min and were subsequently washed with 750 μl of 0,5 % AcOH. The cartridges were dried under vaccum for 5 min.
Elution: Temozolomide was desorbed with 1,25 ml of MeOH.
Evaporation: The eluted solutions were evaporated under a stream of nitrogen at room temperature and reconstituted in 200 μl of 0,5% AcOH.
Authors: Shen F Decosterd L A Gander M Leyvraz S Biollaz J Lejeune F
Source: Journal of Chromatography B 667 (1995) 291-300
Details
Legend: Nominell concentration (µg/ml) Recovery (mean +/- S.D.) (%)
Recovery rates
The absolute recovery of domperidone from human plasma using this method was
83+/- 2,17%. The mean recovery of 0,2